Sensitivity of an infrared gas analyzer used in the differential mode, to partial gas pressures of carbon dioxide and water vapor in the bulk air
Year:
1981
Source of publication :
Agricultural Meteorology
Authors :
אנוך, צבי
;
.
כהן, יחזקאל
;
.
Volume :
24
Co-Authors:
Enoch, H.Z., Division of Agricultural Meteorology, Agricultural Research Organization, The Volcani Center, Bet Dagan, Israel
Cohen, Y., Division of Agricultural Meteorology, Agricultural Research Organization, The Volcani Center, Bet Dagan, Israel
Facilitators :
From page:
131
To page:
138
(
Total pages:
8
)
Abstract:
An improved procedure for the preparation of calibration gases was used to determine the electrical output per unit specific gas pressure (the sensitivity) of carbon dioxide and water vapor infrared gas analyzers (IRGA) used in the differential mode. The IRGA instruments' sensitivity to differential carbon dioxide pressure decreased linearly as the partial carbon dioxide pressure increased. The analog linear decrease in sensitivity was also shown for a water vapor IRGA used in the differential mode. p] Near plant communities there is a djurnal variation in partial carbon dioxide and water vapor pressures which often reaches 20P and 1kP, respectively. Increases of 1P in partial carbon dioxide pressure and of 1 kP in partial water vapor pressure, reduce the IRGA sensitivity by 1 and 19%, respectively. p] If these temporal changes in IRGA sensitivity are ignored, an error of up to 20% in the estimation of carbon dioxide and water vapor gradients above plant communities can be encountered. A procedure is suggested for correcting measurements of water vapor and carbon dioxide gradients for such errors. © 1981.
Note:
Related Files :
עוד תגיות
תוכן קשור
More details
DOI :
10.1016/0002-1571(81)90039-X
Article number:
Affiliations:
Database:
סקופוס
Publication Type:
מאמר
;
.
Language:
אנגלית
Editors' remarks:
ID:
23822
Last updated date:
02/03/2022 17:27
Creation date:
17/04/2018 00:02
You may also be interested in
Scientific Publication
Sensitivity of an infrared gas analyzer used in the differential mode, to partial gas pressures of carbon dioxide and water vapor in the bulk air
24
Enoch, H.Z., Division of Agricultural Meteorology, Agricultural Research Organization, The Volcani Center, Bet Dagan, Israel
Cohen, Y., Division of Agricultural Meteorology, Agricultural Research Organization, The Volcani Center, Bet Dagan, Israel
Sensitivity of an infrared gas analyzer used in the differential mode, to partial gas pressures of carbon dioxide and water vapor in the bulk air
An improved procedure for the preparation of calibration gases was used to determine the electrical output per unit specific gas pressure (the sensitivity) of carbon dioxide and water vapor infrared gas analyzers (IRGA) used in the differential mode. The IRGA instruments' sensitivity to differential carbon dioxide pressure decreased linearly as the partial carbon dioxide pressure increased. The analog linear decrease in sensitivity was also shown for a water vapor IRGA used in the differential mode. p] Near plant communities there is a djurnal variation in partial carbon dioxide and water vapor pressures which often reaches 20P and 1kP, respectively. Increases of 1P in partial carbon dioxide pressure and of 1 kP in partial water vapor pressure, reduce the IRGA sensitivity by 1 and 19%, respectively. p] If these temporal changes in IRGA sensitivity are ignored, an error of up to 20% in the estimation of carbon dioxide and water vapor gradients above plant communities can be encountered. A procedure is suggested for correcting measurements of water vapor and carbon dioxide gradients for such errors. © 1981.
Scientific Publication
You may also be interested in