Low-Cost Dispersive Hyperspectral Sampling Scanner for Agricultural Imaging Spectrometry
Year:
2023
Source of publication :
IEEE Sensors Journal
Authors :
Afgin, Yiftah
;
.
Bonfil, David J.
;
.
Brikman, Roman
;
.
El-Atrash, Saker
;
.
Klapp, Iftach
;
.
Reshef, Liad
;
.
Tsror, Leah
;
.
Volume :
Co-Authors:

Or Arad; 
Loran Cheplanov; 
Yiftah Afgin; 
Liad Reshef; 
Roman Brikman; 
Saker Elatrash; 
Adrian Stern; 
Leah Tsror; 
David J. Bonfil;
Iftach Klapp;
 

Facilitators :
From page:
0
To page:
0
(
Total pages:
1
)
Abstract:

A hyperspectral (HS) imaging system in the 400–1650 nm range was developed based on a point spectrometer matched to a double-wedge prism scanner (Risley prism). This type of scanner is efficient and robust in monochromatic laser systems operating under field conditions. Here, we adopted this scanning principle to develop an inexpensive replacement for the HS camera, suitable for application in precision agriculture and environmental monitoring. However, the application of such a scanner in HS imaging is challenging, due to the broadband nature of HS imaging and the dispersive nature of prisms. The sampling characteristics exhibit wavelength-dependent and space-variant behavior, which affect both ground pixel size and location. In this work, we modeled light propagation through the proposed system using a ray-tracing approximation. Based on this model, we proposed a three-step calibration process. A prototype was built and optimized, followed by an intensive validation process that showed a high correlation between the scanner’s performance and model prediction of ground sample diameter, as well as accuracy in the spectral measurement of the system. Testing reflectance measurement accuracy showed a normalized sum of absolute difference (NSAD) of 4.24% and less. The calibration process was demonstrated in simulations. Finally, preliminary field experiments with two spectrometers in the range of 400–899 nm (Vis–NIR), and 950–1650 nm shortwave infrared (SWIR) showed that the proposed system can support outdoor monitoring, thus having the potential capabilities for supporting agricultural monitoring tasks.

Note:
Related Files :
Hyperspectral Sampling
Imaging spectroscopy (IS)
Show More
Related Content
More details
DOI :
10.1109/JSEN.2023.3282835
Article number:
0
Affiliations:
Database:
Google Scholar
Publication Type:
article
;
.
Language:
English
Editors' remarks:
ID:
66439
Last updated date:
05/11/2023 18:43
Creation date:
05/11/2023 18:42
You may also be interested in
Scientific Publication
Low-Cost Dispersive Hyperspectral Sampling Scanner for Agricultural Imaging Spectrometry

Or Arad; 
Loran Cheplanov; 
Yiftah Afgin; 
Liad Reshef; 
Roman Brikman; 
Saker Elatrash; 
Adrian Stern; 
Leah Tsror; 
David J. Bonfil;
Iftach Klapp;
 

Low-Cost Dispersive Hyperspectral Sampling Scanner for Agricultural Imaging Spectrometry

A hyperspectral (HS) imaging system in the 400–1650 nm range was developed based on a point spectrometer matched to a double-wedge prism scanner (Risley prism). This type of scanner is efficient and robust in monochromatic laser systems operating under field conditions. Here, we adopted this scanning principle to develop an inexpensive replacement for the HS camera, suitable for application in precision agriculture and environmental monitoring. However, the application of such a scanner in HS imaging is challenging, due to the broadband nature of HS imaging and the dispersive nature of prisms. The sampling characteristics exhibit wavelength-dependent and space-variant behavior, which affect both ground pixel size and location. In this work, we modeled light propagation through the proposed system using a ray-tracing approximation. Based on this model, we proposed a three-step calibration process. A prototype was built and optimized, followed by an intensive validation process that showed a high correlation between the scanner’s performance and model prediction of ground sample diameter, as well as accuracy in the spectral measurement of the system. Testing reflectance measurement accuracy showed a normalized sum of absolute difference (NSAD) of 4.24% and less. The calibration process was demonstrated in simulations. Finally, preliminary field experiments with two spectrometers in the range of 400–899 nm (Vis–NIR), and 950–1650 nm shortwave infrared (SWIR) showed that the proposed system can support outdoor monitoring, thus having the potential capabilities for supporting agricultural monitoring tasks.

Scientific Publication
You may also be interested in